Scaling liquid biopsy biomarker discovery: Multi-center, analytical validation of high resolution ctDNA profiling using TSO500 v2 on NovaSeq X Plus

海报缩略图:Scaling liquid biopsy biomarker discovery: Multi-center, analytical validation of high resolution ctDNA profiling using TSO500 v2 on NovaSeq X Plus
编号 3245 展板 10 时间 4/20 02:00–05:00 区域 Section 22 主讲 Bart Tegenbos
分会场 Genomic Profiling to Understand Cancer Biology
查看完整资料 下载 PDF 登录后可访问当前开放资料 AACR 官方页面 ↗

作者与单位

Gabriela Edwards Faret1, Emmanuel Rivière1, Silvie Franck1, Bart Tegenbos1, Lisa Van den Bossche1, Joke Verbist1, Lien Heyrman1, Magdalena Lesnicki2, Rachel Raz2, Brian Chapman2, Eric Harness2, Jan Van de Velde1, Dirk Goossens1, Jurgen Del Favero1

1CellCarta, Antwerp, Belgium,2CellCarta, Naperville, IL

摘要 Abstract

Illumina's TruSight™ Oncology 500 (TSO500) ctDNA v2 assay, enables comprehensive genomic profiling from plasma, allowing non-invasive, repeatable monitoring of tumor evolution during clinical studies. The assay covers a broad panel of 523 clinically relevant cancer genes and provides sensitive detection of key biomarkers such as SNVs, indels, CNVs, gene fusions, MSI, and TMB, all from a simple blood draw. Therefore, this RUO assay is well-suited for exploratory biomarker discovery, response monitoring, and mechanistic insights during immunotherapy and targeted therapy development without the need for fresh tissue biopsies. A validation study was performed to characterize the analytical performance of TSO500 ctDNA v2 assay on the NovaSeq™ X Plus platform for comprehensive genomic profiling of 523 cancer-related genes in ctDNA. The assay was evaluated at two CellCarta laboratories (Antwerp, Belgium and Naperville, USA). Accuracy was assessed using reference materials with a variant allele frequency (VAF) threshold of 0.5% for SNVs and indels. Across both sites, the assay demonstrated >95% overall percent agreement for SNV, indel, CNV, MSI, and TMB results, with positive percent agreement >93.5% and negative percent agreement >95% for SNV, indel, and CNV (no negative reference materials were available for MSI and TMB). Precision assessments showed >97% concordance for intra-run, inter-run, and inter-operator comparisons across all variant classes. Limit of detection verification confirmed 100% detection of SNVs down to 0.25% VAF, indels down to 0.5% VAF, and CNVs at >1.0 fold change. The optimal input for the assay was determined to be 20 ng ctDNA; however, evaluation of lower inputs (10 ng and 5 ng) demonstrated that while these samples did not consistently meet QC thresholds, SNV and indel detection remained robust, with >98% concordance for SNVs and >95% for indels. Lower input material may therefore be used for exploratory analyses, with the understanding that reduced coverage may limit detection sensitivity (potential false negatives) but does not introduce false-positive variant calls. These results confirm that the TSO500 ctDNA v2 assay delivers reproducible, high-resolution genomic profiling across sites, supporting its applicability for exploratory biomarker analyses and longitudinal response monitoring in clinical studies.
利益披露 Disclosure
G. Edwards Faret, None.. E. Rivière, None.. S. Franck, None.. B. Tegenbos, None.. L. Van den Bossche, None.. J. Verbist, None.. L. Heyrman, None.. M. Lesnicki, None.. R. Raz, None.. B. Chapman, None.. E. Harness, None.. J. Van de Velde, None.. D. Goossens, None.. J. Del Favero, None.

在会议检索中打开